Field(s) of Calibration

Electrical (dc and low frequency), pressure and temperature

SCC Accreditation: (ISO/IEC 17025)

  • Accredited Laboratory 810
  • First issued
  • Issue 1.0e

This scope of calibration capabilities is published by the CLAS program of the National Research Council of Canada (NRC) in close co-operation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform the listed calibrations at the given Calibration Measurement Capability (see Supplementary note C and note D) with traceability to the International System of Units (SI) or to standards acceptable to the CLAS program.

Measured Quantity & Range or Instrument Frequency Calibration Measurement Capability expressed as an Uncertainty (±)
(See supplementary notes)
Type Remarks
Frequency
1 Hz to 10 MHz 0.02 ppm + 2 uHz II Measure / Source using a GPS Frequency standard
Voltage, DC
0 mV to 220 mV 0.0008 % + 0.6 μV II Source. For the calibration of dc voltage measuring devices.
220 mV to 2.2 V 0.0007 % + 1 μV
2.2 mV to 11 V 0.0007 % + 3.5 μV
11 mV to 22 V 0.0007 % + 6.5 μV
22 V to 220 V 0.0008 % + 80 μV
220 V to 1100 V 0.0009 % + 500 μV
0 V to 100 mV 0.0009 % + 0.3 μV Measure. For the calibration of dc voltage generating devices.
100 mV to 1 V 0.0008 % + 0.3 μV
1 V to 10 V 0.0008 % + 0.5 μV
10 V to 100 V 0.0010 % + 30 μV
100 V to 1000 V 0.0010 % + 100 μV
Voltage, AC
220 μA to 1100 V 10 Hz to 100 kHz 0.0078 % to 1.3 % II Source. For the calibration of sinewave voltage measurement devices. See Annex A for details.
1 mV to 700 V 10 Hz to 100 kHz 0.007 % to 1.5 % Measure. For the calibration of voltage sources using a digital multimeter. See Annex C for details.
Current, DC
0 to 220 μA 0.005 % + 8 nA II Source. For the calibration of dc current measuring devices.
220 μA to 2.2 mA 0.005 % + 8 nA
2.2 mA to 22 mA 0.005 % + 80 nA
22 mA to 220 mA 0.006 % + 0.8 μA
220 mA to 2.2 A 0.008 % + 25 μA
2.2 A to 11 A 0.036 % + 480 μA
0 to 100 nA 0.003 % + 40 pA Measure. For the calibration of dc current generating devices using a digital multimeter.
100 nA to 1 μA 0.002 % + 40 pA
1 μA to 10 μA 0.002 % + 0.1 nA
10 μA to 100 μA 0.002 % + 0.8 nA
100 μA to 1 mA 0.002 % + 5 nA
1 mA to 10 mA 0.002 % + 50 nA
10 mA to 100 mA 0.0035 % + 0.5 μA
100 mA to 1 A 0.011 % + 10 μA
Current, AC
10 μA to 11 A 10 Hz to 10 kHz 0.02 % to 7 % II Source. For the calibration of sinewave current measurement devices. See Annex B for details.
10 μA to 1 A 10 Hz to 10 kHz 0.03 % to 1 % Measure. For the calibration of current sources using a digital multimeter. See Annex D for details.
Measured Quantity & Range or Instrument Calibration Measurement Capability expressed as an Uncertainty (±)
(See supplementary notes)
Type Remarks
Resistance
0 Ω 50 μΩ II Source. For the calibration of resistance measurement devices.
1 Ω 95 μΩ/Ω
1.9 Ω 95 μΩ/Ω
10 Ω 28 μΩ/Ω
19 Ω 27 μΩ/Ω
100 Ω 17 μΩ/Ω
190 Ω 17 μΩ/Ω
1 kΩ 13 μΩ/Ω
1.9 kΩ 13 μΩ/Ω
10 kΩ 12 μΩ/Ω
19 kΩ 12 μΩ/Ω
100 kΩ 14 μΩ/Ω
190 kΩ 14 μΩ/Ω
1 MΩ 20 μΩ/Ω
1.9 MΩ 21 μΩ/Ω
10 MΩ 40 μΩ/Ω
19 MΩ 47 μΩ/Ω
100 kΩ 110 μΩ/Ω
0.47 Ω to 10 Ω 0.0015 % + 50 μΩ Measure capability using a digital multimeter.
10 Ω to 100 Ω 0.0012 % + 0.5 mΩ
100 Ω to 1 kΩ 0.001 % + 0.5 mΩ
1 kΩ to 10 kΩ 0.001 % + 5 mΩ
10 kΩ to 100 kΩ 0.001 % + 50 mΩ
100 kΩ to 1 mΩ 0.0015 % + 2 Ω
1 mΩ to 10 mΩ 0.005 % + 100 Ω
10 mΩ to 100 mΩ 0.05 % + 1 kΩ
100 mΩ to 1GΩ 0.5 % + 10 kΩ
Pressure
Hydraulic pressure measurement
1.3 to 137.8 MPa 0.009 % + 130 Pa II For the calibration of pressure measuring devices by comparison against primary standard pressure balance DH5304
200 to 20000 psi 0.009 % + 0.022 psi
Low pressure measurement
0 to 750 Pa 0.009 % + 0.064 Pa II For the calibration of pressure measuring devices by comparison against low pressure controller Ruska 7250 LP
0 to 3 inH2 0.009 % + 0.00026 inH2O
750 to 3750 Pa 0.010 % + 0.051 Pa
3 to 15.1 inH2O 0.010 % + 0.00020 inH2O
3750 to 7500 Pa 0.010 % + 0.038 Pa
15.1 to 30 inH2O 0.010 % + 0.00015 inH2O
Gauge pressure measurement
0 to 500 kPa 0.0075 % + 7 Pa II For the calibration of pressure measuring devices by comparison against digital piston gauge DH DPG10G
0 to 72.5 psi 0.0075 % + 0.0010 psi
0 to 2000 kPa 0.0075 % + 24 Pa
0 to 290 psi 0.0075 % + 0.0035 psi
0 to 5000 kPa 0.0075 % + 58 Pa
0 to 725 psi 0.0075 % + 0.0084 psi
0 to 17236 kPa 0.0075 % + 220 Pa
0 to 2500 psi 0.0075 % + 0.032 psi
Gauge pressure measurement
1.4 to 172 kPa 0.0055 % + 0.07 Pa II For the calibration of pressure measuring devices by comparison against primary piston gauge Ruska 2465
0.2 to 25 psi 0.0055 % + 0.00001 psi
12 to 700 kPa 0.0055 % + 1.4 Pa
1.7 to 100 psi 0.0055 % + 0.0002 psi
14 to 3500 kPa 0.0055 % + 4 Pa
2 to 500 psi 0.0055 % + 0.0006 psi
14 to 7000 kPa 0.0055 % + 8 Pa
2 to 1000 psi 0.0055 % + 0.0012 psi
Absolute pressure measurement
Pneumatic
1.4 to 172 kPa a 0.0055 % + 2.7 Pa II For the calibration of pressure measuring devices by comparison Direct comparison against primary piston gauge (Ruska 2465).
0.2 to 25 psi a 0.0055 % + 0.0004 psi
12 to 700 kPa a 0.0055 % + 4 Pa
1.7 to 100 psi a 0.0055 % + 0.0006 psi
14 to 3500 kPa a 0.0055 % + 5 Pa
2 to 500 psi a 0.0055 % + 0.000725 psi
14 to 7000 kPa a 0.0055 % + 9 Pa
2 to 1000 psi a 0.0055 % + 0.0013 psi
Electrical Calibration of Temperature Indicators and Simulators
RTD source and measure
−200 °C to 850 °C 0.0034 °C to 0.026 °C II For the calibration of Pt100 (375), Pt100 (385), Pt100 (389), Pt100 (391), Pt100 (3926) RTD indicators and simulators by electrical simulation of temperature.
Thermocouple source and measure
−270 °C to 1768 °C 0.13 °C to 0.7 °C II For the calibration of type K, J, E, T, N, R, S, B thermocouple indicators and simulators by electrical simulation of temperature.
Temperature
−20 °C to 140 °C 0.06 °C II For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a liquid bath.
140 °C to 650 °C 0.3 °C For the calibration of RTD with or without temperature indicators by comparison against the laboratory temperature standard in a dry block.

Annex A

Best Uncertainty, ± {(ppm of sourced ac voltage) + residual}, of AC voltage generation for the calibration of voltage measuring devices
Voltage, AC Frequency
10 Hz to
20 Hz
20 Hz to
40 Hz
40 Hz to
20 kHz
20 kHz to
50 kHz
50 kHz to
100 kHz
100 kHz to
300 kHz
300 kHz to
500 kHz
500 kHz to
1 MHz
220 μV to 2.2 mV 550 + 4.5 μV 210 + 4.5 μV 105 + 4.5 μV 370 + 5 μV 850 + 7 μV 1100 + 13 μV 1700 + 25 μV 3400 + 25 μV
2.2 mV to 22 mV 550 + 5 μV 210 + 5 μV 105 + 5 μV 370 + 5 μV 850 + 7 μV 1100 + 12 μV 1700 + 25 μV 3400 + 25 μV
22 mV to 220 mV 550 + 13 μV 210 + 8 μV 105 + 8 μV 320 + 8 μV 850 + 25 μV 1100 + 25 μV 1700 + 25 μV 3400 + 80 μV
220 mV to 2.2 V 500 + 80 μV 160 + 25 μV 75 + 6 μV 120 + 16 μV 250 + 70 μV 430 + 130 μV 1050 + 350 μV 2200 + 850 μV
2.2 V to 22 V 500 + 800 μV 160 + 25 μV 75 + 60 μV 120 + 160 μV 250 + 350 μV 500 + 1500 μV 1250 + 4300 μV 2700 + 8500 μV
22 V to 220 V 500 + 8 mV 160 + 2.5 mV 80 + 0.8 mV 220 + 3.5 mV 500 + 8 mV 1500 + 90 mV 4700 + 90 mV 11,500 + 190 mV
Best Uncertainty, ± {(ppm of sourced ac voltage) + residual}, of AC voltage generation for the calibration of voltage measuring devices
Voltage, AC Frequency
40 Hz to
1 kHz
1 kHz to
20 kHz
20 kHz to
30 kHz
30 kHz to
50 kHz
50 kHz to
100 kHz
220 V to 1100 V 90 + 4 mV 165 + 6 mV 600 + 11 mV N//A N//A
220 V to 750 V N//A N//A N//A 600 + 11 mV 2300 + 45 mV

Annex B

Best Uncertainty, ± {(ppm of sourced ac current) + residual}, of AC current generation for the calibration of voltage measuring devices
Current, AC Frequency
10 Hz to
20 Hz
20 Hz to
40 Hz
40 Hz to
1 kHz
1 kHz to
5 kHz
5 kHz to
10 kHz
10 μA to 220 μA 700 + 25 nA 350 + 20 nA 140 + 16 nA 600 + 40 nA 1600 + 80 nA
220 μA to 2.2 mA 700 + 40 nA 350 + 35 nA 140 + 35 nA 600 + 400 nA 1600 + 800 nA
2.2 mA to 22 mA 700 + 400 nA 350 + 350 nA 140 + 350 nA 600 + 4000 nA 1600 + 8,000 nA
22 mA to 220 mA 700 + 4 μA 350 + 3.5 μA 140 + 3.5 μA 600 + 40 μA 1600 + 80 μA
220 mA to 2.2 A N/A 650 + 35 μA 650 + 35 μA 750 + 80 μA 8,500 + 160 μA
2.2 V to 11 A N/A N/A 460 + 170 μA 950 + 380 μA 3600 + 750 μA

Annex C

Best Uncertainty, ± {(% of measured AC voltage) + residual}, of AC voltage measurement for the calibration of voltage generating devices
Voltage, AC Frequency
1 Hz to
40 Hz
40 Hz to
1 kHz
1 kHz to
20 kHz
20 kHz to
50 kHz
50 kHz to
100 kHz
100 kHz to
300 kHz
1 mV to 10 mV 0.03 + 3 μV 0.02 + 1.1 μV 0.03 + 1.1 μV 0.1 + 1.1 μV 0.5 + 1.1 μV 4 + 2 μV
10 mV to 100 mV 0.007 + 4 μV 0.007 + 2 μV 0.014 + 2 μV 0.03 + 2 μV 0.08 + 2 μV 0.3 + 10 μV
100 mV to 1 V 0.007 + 40 μV 0.007 + 20 μV 0.014 + 20 μV 0.03 + 20 μV 0.08 + 20 μV 0.3 + 100 μV
1 V to 10 V 0.007 + 400 μV 0.007 + 200 μV 0.014 + 200 μV 0.03 + 200 μV 0.08 + 200 μV 0.3 + 1000 μV
10 V to 100 V 0.02 + 4000 μV 0.02 + 2000 μV 0.02 + 2000 μV 0.035 + 2000 μV 0.12 + 2000 μV 0.4 + 10000 μV
100 V to 700 V 0.04 + 40000 μV 0.04 + 20000 μV 0.06 + 20000 μV 0.12 + 20000 μV 0.3 + 20000 μV
Best Uncertainty, ± {(% of measured AC voltage) + residual}, of AC voltage measurement for the calibration of voltage generating devices
Voltage, AC Frequency
300 kHz to
1 MHz
1 MHz to
2 MHz
2 MHz to
4 MHz
4 MHz to
8 MHz
8 MHz to
10 MHz
1 mV to 10 mV 1.2 + 5 μV 7 + 7 μV 7 + 7 μV 20 + 8 μV
10 mV to 100 mV 1 + 10 μV 1.5 + 10 μV 4 + 70 μV 4 + 80 μV 15 + 100 μV
100 mV to 1 V 1 + 100 μV 1.5 + 100 μV 4 + 700 μV 4 + 800 μV 15 + 1000 μV
1 V to 10 V 1 + 1000 μV 1.5 + 1000 μV 4 + 7000 μV 4 + 8000 μV 15 + 10000 μV
10 V to 100 V 1.5 + 10000 μV

Annex D

Best Uncertainty, ± {(% of measured AC current) + residual}, of AC current measurement for the calibration of current generating devices
Current, AC Frequency
10 Hz to
20 Hz
20 Hz to
45 Hz
45 Hz to
100 Hz
100 Hz to
5 kHz
5 kHz to
20 kHz
20 kHz to
50 kHz
50 kHz to
100 kHz
10 μA to 100 μA 0.4 + 0.03 μA 0.15 + 0.03 μA 0.06 + 0.03 μA 0.06 + 0.03 μA
100 μA to 1 mA 0.4 + 0.2 μA 0.15 + 0.2 μA 0.06 + 0.2 μA 0.03 + 0.2 μA 0.06 + 0.2 μA 0.4 + 0.4 μA 0.55 + 1.5 μA
1 mA to 10 mA 0.4 + 2 μA 0.15 + 2 μA 0.06 + 2 μA 0.03 + 2 μA 0.06 + 2 μA 0.4 + 4 μA 0.55 + 15 μA
10 mA to 100 mA 0.4 + 20 μA 0.15 + 20 μA 0.06 + 20 μA 0.03 + 20 μA 0.06 + 20 μA 0.4 + 40 μA 0.55 + 150 μA
100 mA to 1 A 0.4 + 200 μA 0.16 + 200 μA 0.08 + 200 μA 0.1 + 200 μA 0.3 + 200 μA 1 + 400 μA