CLAS Certificate Number 2016-01

La Cie J. Chevrier Instruments Inc.

4850, bd Gouin est
Montréal-Nord, Québec H1G 1A2

Contact

Denis Chevrier
Tel.: 514-328-2550 poste 22
Toll free: 1-800-522-1226
Fax: 514-327-0604
Emailinfo@chevrierinstruments.com

Clients served

  • All interested parties

Fields of calibration

  • Mechanical
  • Temperature

SCC accreditation (ISO/IEC 17025)

  • Accredited Laboratory no. 822
  • First issued 2016-02-23
  • Issue 1.1 2017-02-06

This scope of calibration capabilities is published by the CLAS program of the National Research Council of Canada (NRC) in close co-operation with the laboratory accreditation program of the Standards Council of Canada (SCC), Canada's accreditation body for calibration and testing laboratories. The SCC accredits the capability of the named laboratory for being able to perform the listed calibrations at the given Calibration Measurement Capability (see Supplementary note C and note D) with traceability to the International System of Units (SI) or to standards acceptable to the CLAS program.

Measured Quantity & Range or Instrument Calibration and Measurement Capability expressed as an Uncertainty (±)
(See supplementary notes)
Type of Service Remarks
Mechanical
Pressure indicators
0 psi to 1000 psi
0.4 psi II

For the calibration of Analog gauges, Digital gauges, Pressure transducers, Pressure calibrators

Comparison method with digital pressure calibrator

1000 psi to 3000 psi
1 psi
3 psi to 16000 psi
2 to 6 psi

For the calibration of Analog gauges, Digital gauges, Pressure transducers, Pressure calibrators

Hydraulic deadweight tester

Differential or static pressure
0 KPa to 18 KPa
0.003 KPa II

Digital pressure calibrator

 

18 KPa to 200 kPa
0.03 KPa
Temperature
Resistance temperature devices
0.01 °C, triple point of water
0.01 °C II

For the calibration of Resistance Temperature Devices (PRT, RTD) using:
Triple point of water cell

29.7646 °C, melting point of gallium
0.006 °C

For the calibration of Resistance Temperature Devices (PRT, RTD) using:
Gallium melting point cell